Fully Edentulous CBCT Scan: Dual Scan Protocol of Maxilla

1.  On a dry, well-fitting denture, place 6 x-ray markers: 

  • Place 3 markers on the facial/buccal surface and an additional 3 markers on the palatal surface
  • Place the markers on varying axial planes.
  • Example: if you place a marker on the buccal area of #2-3, the place your next marker on the lingual area of #4-5 area, the next marker on the facial of #8-9 area and so on.

*Do not place the markers on the intaglio surface of the denture.*

2.  Pre-Scan Patient Instructions:

  • Use a bite registration for separation, secure fit, and to eliminate rocking
  • Confirm fit by asking the patient “Is the denture fitting correctly?”
  • Breathe through your nose, close your eyes, and place the tongue at the back of the mouth

3.  First Scan: Capture CBCT of ONLY the maxilla denture with the x-ray markers

  • Position appliance on a foam block
  • Orientate teeth occlusal DOWN

CBCT General Settings: 6cm – 8cm, 0.3 – 0.4 Voxel

4.  Second Scan: Capture CBCT of the patient WEARING maxilla denture with x-ray markers.

Vital Anatomy to be captured:

  • Superior: Mid Sinus / below orbit
  • Inferior: Capture 1-3 mm of mandible dentition

CBCT General Settings: Maxl 6cm – 8cm, 0.3 – 0.4 Voxel

5.  When scanning a patient following Dual Scan Protocol:

  • Review the scan carefully to ensure there is no “Black Air Space” between the denture and the soft tissue of the patient

red-x15Rejected Scan  

WRONG1

green checkAccepted Scan  

RIGHT1

6.  Export and Upload CBCT Scans to Implant Concierge

  • Export the CBCT scans individually, in a multi-file DICOM format to the desktop
  • Right-click on each folder, go to “Send To”, and select “Compressed (Zipped) Folder”
  • Upload both zipped/compressed folders to Implant Concierge